<?xml version="1.0" standalone="yes"?>
<?xml-stylesheet type="text/xsl" href="..\..\XSLTemplates\ErrorsStyleSheet.xsl"?>
<Error>
<ErrorNumber>17737</ErrorNumber>
<ErrorFamily>Exposure Head</ErrorFamily>
<ErrorMessage>
HEAD: case temperature low alarm, &lt;temperature&gt; C
</ErrorMessage>
<AppliesTo>
<UnnumberedList>
<ListItem><Text>Thermal</Text></ListItem>
</UnnumberedList>
</AppliesTo>
<ServiceInformation>
<DescriptionText><text>
Every thirty seconds, the thermal head firmware reads the case temperature and
compares the reading to a threshhold.
</text></DescriptionText>
<DescriptionText><text>
For both air-cooled and water-cooled thermal heads, the threshhold is hardcoded
to 10 degrees C.  If the threshhold has been exceeded, the thermal head is
shutdown and any plot in progress is aborted.
</text></DescriptionText>
<DescriptionText><text>
This error code applies to Thermal Head firmware version 1.30 and above.
</text></DescriptionText>
</ServiceInformation>
<OperatorAction>
<DescriptionText><text>
Call service.
</text></DescriptionText>
</OperatorAction>
<Cause>
Verify the temperature and humidity of the environment local to the machine.
</Cause>
<Cause>
Check the thermal head cooling ( thermal heads are air-cooled or water-cooled ).
Is the water-cooling functioning?
</Cause>
<SeeAlso>
<UnnumberedList>
<Link><LinkAddress>../nvs/head_cal_casehigh.xml</LinkAddress><LinkTitle>NVS: head casehigh</LinkTitle></Link>
<Link><LinkAddress>../nvs/head_cal_caselow.xml</LinkAddress><LinkTitle>NVS: head caselow</LinkTitle></Link>
</UnnumberedList>
</SeeAlso>
<OtherLinks>
<Link><LinkAddress>..\..\ShowAllErrors.htm</LinkAddress><LinkTitle>Firmware 0 Error Index</LinkTitle></Link>
<Link><LinkAddress>..\..\ShowAllCommands.htm</LinkAddress><LinkTitle>Firmware 0 Command Index</LinkTitle></Link>
<Link><LinkAddress>..\..\ShowAllNVS.htm</LinkAddress><LinkTitle>Firmware 0 NVS Index</LinkTitle></Link>
</OtherLinks>
<Updated>
Updated: Wed Jul 16 19:33:45 2008 PST
</Updated>
<Links>
</Links>
</Error>